Tumor Lysate Nitrocellulose Print 9-27-12

2015-01-13

azim58 - Tumor Lysate Nitrocellulose Print 9-27-12


The print will actually take place on 9-28-12

Print onto Grace Biolabs Nitrocellulose Slides (dimensions are on that
page)

The pad starts about 3 mm down and 2.5 mm over.
The pad covers about 60 mm down and 20 mm over.
The print positions will start as follows


Plate mixing before printing at 2000 rpm 15 s. 3000 rpm 10 min centrifuge

Pin Operation

o sonicate 30 s
o wash in 10% ethanol 10 s (ethanol is better for the tubes in the
instrument)
o dry 5 s

There will be a double wash inbetween subarray or new slide

Note that each well will have it's own "dip"

Distance of spots: 1-1.5 mm

No landing lights used this time, but if they were, here is the
information for that:
Landing Light Concentration: 1:100
Landing Light Identity: a biotinylated peptide. Therefore, in order to
detect the landing lights, it is necessary to use streptavidin
AlexaFluor. If direct labeled secondary is used, the landing lights will
not be visible unless a small of amount (such as 0.1 nM) of streptavidin
AlexaFluor is added as well.