Cell Lysate Print 14
2015-01-13azim58 - Cell Lysate Print 14
Cell Lysate Print 14 (Grace BioLabs Nitrocellulose Slides; blocking
tested with Super G and BSA; primary incubation time tested with 1 hr 37
and 16 hr 23 C)
found at 2-11-12
"F:\kurt\storage\CIM Research Folder\DR\2013\9-5-13\lab notes
corresponding to cell lysate print 14 experiment 2-11-12.pdf"
Since the last overnight print run did not work well, we will test
another print run 1 hr and overnight with the different blocking buffers.
Setup:
1 slide BSA primary incubation 1 hr
1 slide Super G primary incubation 1 hr
1 slide BSA primary incubation 16 hr
1 slide Super G primary incubation 16 hr
Slide print order in printing machine: the slide closest to the open room
was printed first and the other slides were printed in order working back.
Johnny said that there was a slight mistake and reprint of about the
first 3 spots on the top part of the slides. Hopefully, this will not
cause too much of a problem.
Raw Data can be found here:
C:\kurt\storage\CIM Research Folder\DR\2012\8-25-12\8-25-12_1140\Printing
Library onto array\Cell Lysate Test Prints\blocking and time test 2-11-12
old link
"L:\storage\CIM Research Folder\kwhittem\Records in CIM
Folder\Categorical Records\Biodesign\Immunosignature to Antigen Bridge
Project\Tumor cDNA Library\Printing Library onto array\Cell Lysate Test
Prints\blocking and time test 2-11-12"
Summary
- PDF Summary
Folder\DR\2013\6-22-13\wiki_download\comparison of blocking buffers
and primary incubation time 2-22-12 1.svg"
o basically The Super G results in less smearing, and the overnight
conditions resulted in higher signals. I was able to detect 1/1000
and even 1/10000 SMC1fs dilutions above background.
- Original Inkscape Files
2-22-12 1.svg
o "F:\kurt\storage\CIM Research
Folder\DR\2013\6-25-13\wiki_download\comparison of blocking buffers
and primary incubation time 2-22-12 2.svg"
o "F:\kurt\storage\CIM Research
Folder\DR\2013\6-25-13\wiki_download\comparison of blocking buffers
and primary incubation time 2-22-12 3.svg"
see also: How to get good results with lysate on nitrocellulose slides