-Thermocycle Conditions 7-9-13 50 C
2014-08-29azim58 - -Thermocycle Conditions 7-9-13 50 C
Thermocycle Conditions 7-9-13 50 C
has been associated with pm ss tdwn 50 C (pm = primestar max, ss = small
primer start, tdwn = touchdown)
95 C 15 s, 95 C hold, 95 C 2 min, 95 C hold,
98 C 10 s, 60 C 10 s, 72 C 60 s,
98 C 10 s, 59 C 10 s, 72 C 60 s,
98 C 10 s, 58 C 10 s, 72 C 60 s,
98 C 10 s, 57 C 10 s, 72 C 60 s,
98 C 10 s, 56 C 10 s, 72 C 60 s,
98 C 10 s, 55 C 10 s, 72 C 60 s,
98 C 10 s, 54 C 10 s, 72 C 60 s,
98 C 10 s, 53 C 10 s, 72 C 60 s,
98 C 10 s, 52 C 10 s, 72 C 60 s,
98 C 10 s, 51 C 10 s, 72 C 60 s,
(98 C 10 s, 50 C 10 s, 72 C 60 s)X15
4 C hold, 98 C hold
(98 C 10 s, 50 C 10 s, 72 C 60 s)X20
4 C hold
(98 C 10 s, 50 C 10 s, 72 C 60 s)X15
4 C hold
(98 C 10 s, 50 C 10 s, 72 C 60 s)X10
4 C hold
1st part to warm up thermocycler
2nd part to denature
3rd part to add polymerase and denature further
10 decreasing annealing temperature cycles for touchdown
amplification for 15 cycles with the small amount of primer originally
added (1 uL of 2 uM forward and reverse primer each; 0.04 uM final
concentration each primer at this point)
98 C hold for addition of more primer (1 uL of 10 uM forward and reverse
primer each; 0.24 uM final concentration each primer at this point)
20 cycles of amplification
4 C hold for removal of an aliquot if desired (e.g. 20 uL)
15 cycles of amplification
4 C hold for removal of an aliquot if desired (e.g. 20 uL)
10 cycles of amplification
4 C hold for removal of remaining sample (e.g. 10 uL)
Timing
- time from addition of polymerase to addition of more primer:
- time from addition of more primer to removal of 1st aliquot: 80s*20 =
- time from removal of 1st aliquot to removal of 2nd aliquot: 80s*15 =
- time from removal of 2nd aliquot to removal of final aliquot: 80s*10 =
see also PCR by starting with a small amount of primer and later adding
more primer and polymerase portal 4-11-13